Determinatin of chlorogenic acid in hawthorn by HPLC
doi: 10.3969/j.issn.1006-0111.2012.06.018
- Received Date: 2012-06-01
- Rev Recd Date: 2012-09-12
-
Key words:
- chlorogenic acid /
- HPLC /
- hawthorn
Abstract: Objective To establish a HPLC method for determination of chlorogenic acid in the hawthorn. Methods Sunfire C18 (4.6 mm×250 mm, 5 μm) column was used. The mobile phase was methanol:sodium dihydrogen phosphate(25:75). The flow rate was 1.0 ml/min, and the temperature was at 28℃. The detection wavelength was 327 nm. Results The calibration curve was at the range of 7.03~70.3 μg/ml (r=0.999 6), RSD was 0. 92% (n=6) and the average recovery was 96.92%. Conclusion The method was simple, reliable, accurate for determination of chlorogenic acid in the hawthorn.
Citation: | LIANG Rui-shu, ZHANG Chuan, ZOU Hao, JIN Lei. Determinatin of chlorogenic acid in hawthorn by HPLC[J]. Journal of Pharmaceutical Practice and Service, 2012, 30(6): 457-458,477. doi: 10.3969/j.issn.1006-0111.2012.06.018 |