A HPLC analysis method of synthetic ferulic acid
- Received Date: 2001-05-08
-
Key words:
- synthetic ferulic acid /
- HPLC /
- analysis /
Abstract: OBJECTIVE To detect the purity of synthetic ferulic acid.METHODS The synthetic acid was determined by HPLC.A 250mm×4.6mm,C18 Nova-pak column with mobile phase MeOH:H2O(1:1) was used and detected at 254nm.RESULTS The peak area of synthetic ferulic acid had a linear relation with injection mass in range of 0.24~2.0μg,and purity of synthetic product was 99.4%.CONCLUSIONS HPLC was a simple,fast and accurate method for synthetic ferulic acid analysis.
Citation: | ZHANG Xiang-nian, HAN Li-ping, ZHAO Shu-jin. A HPLC analysis method of synthetic ferulic acid[J]. Journal of Pharmaceutical Practice and Service, 2001, (6): 358-359. |