Method verify on quantity of organic solvents remaining
- Received Date: 2003-10-28
-
Key words:
- GC /
- quantity of organic solvents /
- remaining
Abstract: OBJECTIVE To establish a GC method for the determination of organic solvents remaining. METHODS The methods was approved by high efficiency chlomatogram column of INNOWAX(30m×0.32mm×0.25μm),with a detector of FID using direct injection. RESULTS:There is no interfere of every component between ethanol,DMF,acetic acid inside standard and solven. The resolution and sensitivity were acceptable. The detection limit of the three components were 2.03μg/mL, 2.21μg/mL, 3.75μg/mL respectively. CONCLUSIONS This method is accurate to control the organic solvents remaining.
Citation: | HAN Jia-yi, ZHENG Guo-gang. Method verify on quantity of organic solvents remaining[J]. Journal of Pharmaceutical Practice and Service, 2003, (6): 355-357. |